2.4 inch TFT color screen . Resolution: 320*240
Display waveform: 2 channels waveform of ECG, respiration, blood pressure and SpO2 can be both displayed
ECG: With the function of abnormal data store, daily data store and trend display
ECG Cable: 3 leads .10 Leads
NIBP: Manual measurement. Automatically identify the adult and child. System can adjust the maximal pressure for adult and child. The blood pressure waveform can be displayed in real-time
SpO2: Anti-low per-fusion and motion artifact
Trend display: Display the trend of ECG, blood pressure, SpO2, ST segments, premature and respiration in 96 hours
Alarm modes: Visible alarming reminder and speaking voice alarming reminder
Wireless: Bluetooth 4.0
E-Manual: Embedded e-manual for self reading
workable with smart devices (Android &iOS)
ECG
ECG waveform gain: 0.5 mm / mVļ¼1.0 mm / mVļ¼ 2.0 mm / mV
Scanning speed : 7.5mm / s, 15mm / s
Heart rate and alarm range :
Adult: 30 bpm ļ½250 bpm (beats / min)
Precision of ECG: ±5%
Resolution: 1 bpm (Times/Minute)
Sensitivity: ļ¼200uV (Peak-to-Peak Value)
Trigger level threshold: 200nV (Lead II)
Resistance of differential input: ļ¼5MĪ©
Common-mode rejection ratio : ā„ 90 dB
Electrode polarization voltage range: 300mV
Leakage current: ļ¼10uA
Input signal range: ±5mV
ECG abnormal types:Tachycardia, Bradycardia, VPC, Bigeminal rhythm, Trigeminal rhythm, RonT, VTAC, Escape Beat and Sinus arrest.
Alarmļ¼The voice audible alarm and screen visual alarm display dual alarm, under review.
Heart rate alarm rangeļ¼ 30ļ½250 bpm
Alarm Muteļ¼ Less than 10 seconds
Alarm sound pressureļ¼ 45 to 65 db
Frequency response :
Flter : 0.67 Hz ļ½ 40Hz
Guardianship : 0.05 Hzļ½40Hz
Detection: 0.05 Hzļ½100Hz
Calibration signal : 6 mm / mV
ST segment measurement:
Measured range: -2.0mVļ½+2.0mV
Precise: -0.8mVļ½+0.8mV; ±0.02mV or±10%
Renew period: 10s
Pacing pulse suppression :
Range: ± 2mV to 500Mv
Width: 0.1ms ļ½ 2ms
Rise time: 10us ļ½ 100 μs
Inhibition of the pacing pulse detector for fast ECG signal : 2.5V/s
Can inhibit the T-wave amplitude: > 5mV
Noninvasive blood pressure (NIBP)
Measurement: Oscillometric pressure when the blood pressure measurement.
Pressurized automatic calibration zero
Measuring range : 30mmHgļ½280mmHg
Measurement accuracy : ±5mmHgor ±5%
Maximum cuff pressure: Cuff pressure does not exceed 300 mmHg,
The cuff pressure at 15 mmHg or more time is not more than 3min.
Overvoltage protection: Yes
Alarmļ¼ Systolic / diastolic alarm error of ± 2 mmHg
Atrial premature function
Measuring wrist Week 135mm ļ½ 215mm
Oxygen saturation (SpO2)
Measuring range ļ¼ 40% ļ½ 99%
Measurement accuracy: 90% ļ½ 99% Tolerance ± 2%
70% ļ½ 89% Tolerance ± 4%
40% ļ½ 69% Tolerance ± 6%
Alarm range ļ¼ 50% ļ½ 99%
Pulse rate
Measuring range ļ¼ 30 Beats / min ļ½ 80 Beats / min ± 2
80 Beats / min ļ½150 Beats / min ± 4
150 Beats / min ļ½240 Beats / min ,± 5
Measurement accuracyļ¼ Ā± 5%
Resist electric shock ļ¼ CF type
Build-in : 3.7V 1000mh
Power adapterļ¼ 5V/1A 2W battery charger:
Power 2W
Type : Non-AP / APG type Do not prevent inlet device
Work environmentļ¼
Temperature: 5 āļ½40 ā
Humidity: 15% ļ½80%
Atmospheric pressure: 80kPa ļ½105kPa
Storage environmentļ¼
Temperature ļ¼ -20 ā ļ½55ā
Humidity ļ¼ 10%ļ½90%
Atmospheric pressure ļ¼ 50.0 kPa ļ½106.0kPa
Size and Weight
Dimensions ļ¼ 7cm (W) Ć 7cm (H) Ć2.6cm (D) (does not include a wrist strap)
Net Weight ļ¼ 130g
Gross Weight 0.5Kg
Standard:
ECG Cable (3 Leads) …… 1Pcs
SpO2 probe (Soft tip) ….. 1Pcs
Power Adapter ………….. 1Pcs
Cleaning Cloth ……………. 1Pcs
Options:
ECG Cable (10 Leads)
Reviews
There are no reviews yet.